Specifications
| Sensor Type |
Ambient Temp & Humidity |
Thermistor |
K-Thermocouple |
Platinum RTD |
Ambient Temp |
| SKU |
RS010 |
RS020
RS021 |
RS030
RS031 |
RS040 |
RS050 |
| Humidity Accuracy* |
+/-2.0RH from 5.0 to 95.0RH |
|
|
|
|
| Temperature Accuracy* |
+/-0.8F (+/-.44C) from 20.0 to 122.0F +/-1.8F (+/-1.0C) remaing range (range of sensor is -40 to 185F) |
+/-0.9F |
+/-1.8F from -22 to 122F (+/-1.0C from -30 to 50C)
Remainder of the Range: +/- 3.8F |
+/-0.5F from -148.0 to 350.0F (+/-0.3C from -100.0 to 176.0C) |
+/-1.2F from 20.0 to 122.0F (+/-1.0C from -30.0 to 50.0C) |
| Temperature Accuracy (Remainder of Range)* |
|
+/-0.9F |
|
|
|
| Temperature Range |
-40.0 to 185.0F (-40.0 to 85.0C) |
-58.0 to 158.0F (-50.0 to 70.0C) |
-300.0 to 2000.0F (-185.0 to 1093.0C) (Must use A203 for temperatures above 500F) |
-148.0 to 350.0F (-100.0 to 176.0C) |
-22.0 to 122.0F (-30.0 to 50.0C) |
Compatibility
| Sensor Type |
Ambient Temp |
Ambient Temp & Humidity |
Platinum RTD |
Thermistor |
K-Thermocouple |
Dual K-thermocouple |
Dual Thermistor |
| SKU |
RS050 |
RS010 |
RS040 |
RS020 |
RS030 |
RS031 |
RS021 |
| Compatibility |
DSB |
DR8 |
RFL |
|
|
|
|
Platinum RTD Temperature (RS040) Specifications
| Sensor Type |
Ambient Temp & Humidity |
Thermistor |
K-Thermocouple |
Platinum RTD |
| SKU |
RS010 |
RS020
RS021 |
RS030
RS031 |
RS040 |
| Cable Length |
|
|
|
|
| Ambient Operating RH Conditions |
|
|
|
|
| Humidity Accuracy* |
+/-2.0RH from 5.0 to 95.0RH |
|
|
|
| Humidity Range |
|
|
|
|
| Temperature Accuracy* |
+/-0.8F (+/-.44C) from 20.0 to 122.0F +/-1.8F (+/-1.0C) remaing range (range of sensor is -40 to 185F) |
+/-0.9F from -58.0 to 68.0F (+/-0.5C from -50.0 to 20.0C)
+/-1.2F from 69F to 122F |
+/-1.8F from -300.0 to 122.0F (+/-1.0C from -185.0 to 50C) +/-3.8F (+/-2.1C) remaining range |
+/-0.5F from -148.0 to 350.0F (+/-0.3C from -100.0 to 176.0C) |
| Temperature Accuracy (Remainder of Range)* |
|
+/-2.0F over remaining range |
|
|
| Temperature Range |
B9 -40.0 to 185.0F (-40.0 to 85.0C) |
C9 -58.0 to 158.0F (-50.0 to 70.0C) |
-300.0 to 2000.0F (-185.0 to 1093.0C) |
-148.0 to 350.0F (-100.0 to 176.0C) |
*In statements of conformity on calibration certificates, Dickson applies the Simple Acceptance decision rule, as defined in ILAC-G8:09/2019, with a TUR of 4:1 or greater (for Thermocouples, TUR of 2:1 at temperatures -22F° to 122°F), unless otherwise specified and agreed with the customer.